Super-resolution implies that you are imaging below the diffraction (Abbe) limit. While the Fourier ring correlation can be used as a surrogate for resolution (it is frequently used in both cryo-em and fluorescent microscopy) you must define the correlation cutoff or else the measurement is meaningless. I do not see any mention of a cutoff used in the linked paper.
I would be more convinced if they actually imaged something smaller than the diffraction limit of their microscope, like the nuclear pore, which was recently proposed as a standard for the field [2]. Nonetheless I am impressed by the image quality given how un-rigid and prone to thermal drift/expansion their microscope probably is.