by gh02t · 3mo ago · 0 comments
The timing and electronics for doing position sensing of the gamma path through the detector (i.e., what you need to reconstruct an image) are... complex. Dead time and false correlates (e.g. another gamma in the detector at the same time, aka false coincidences) are a thing and they are factors that contribute to uncertainties. There's a bunch of statistical analysis you have to do to try and correct for these effects and pull the signal out of the noise. In a Compton camera, you can correlate the first scatter's trajectory back into a cone of possible incident directions, which you then stack up to form a picture. Hence why I said the image you get out is very blobby and more like a heat map than an image. But you pay a price in efficiency, as you can only count some fraction of the actual interactions in the detector now. I'm a bit rusty on the really nuts and bolts details, I don't work with this type of detector anymore. We mostly try to avoid using this type of detector for our localization, because they are crazy expensive and complex.
Notable sites included most of the hotter mine sites about the globe, Finnish / Russian sea borders, and the odd live underground test here and there (Pokhran-II exchange) .. but enough about that.